Remote Operation
Example
:PG:DATA:PATT:NAME PRBS2^7-1;*OPC?
1
Selects the pattern name for the pattern generator connected to
channel 0 (Jitter) only and uses the *OPC? query command to
determine if the operation is complete.
Command
:PG:DATA:PATTern:NAME?
Description
Return the selected data pattern name.
Example
:PG:DATA:PATT:NAME?
PRBS2^7-1
Queries the selected pattern name of the pattern generator connected
to channel 0 (Jitter) only.
Command
:PG:DATA:PATTern:POLarity
Description
Set the data pattern polarity to INVert, to invert the data pattern
presented at the data outputs, or NONInvert. The default is NONInvert.
Example
:PG:DATA:PATT:POL INV
Sets the pattern polarity to inverted for the pattern generator connected
to channel 0 (Jitter) only.