Keysight Technologies N4960A User Manual page 124

Serial bert 17 and 32 gb/s
Table of Contents

Advertisement

Remote Operation
124
Command
:OUTDelay:COUPling
Description
Set the position of the internal switch to AC (default) or DC. Offset or
termination must be set to 0 V to change coupling to AC or a setting
conflict error message will be generated. If coupling is already set to AC,
it can be changed to DC without any dependencies.
Example
:OUTD:COUP DC
Command
:OUTDelay:COUPling?
Description
Return the position of the internal switch. The returned string is AC or
DC.
Example
:OUTD:COUP?
DC