Command
:ED:DATA:ACCumulation:STOP
Description
Stop an accumulated BER measurement. This command can be issued
regardless of the accumulation stop criteria.
Example
:ED:DATA:ACC:STOP
Command
:ED:DATA:INPut:DELay
Description
Set the sampling delay from –2000 UI to 2000 UI in 0.002 UI
increments. This sets the amount the clock must be shifted to sample
the incoming data to the error detector. In most cases, this value will be
centered in the received eye for best performance using auto alignment.
The optional units are UI (Unit Interval) and mUI (milli Unit Interval). The
default value/unit is 0.00 UI.
The sampling delay can be set manually or automatically by the auto
alignment process. See the :ED:DATA:AALign:EXECute command.
Example
:ED:DATA:INP:DEL 1000UI;*OPC?
1
This command requires some time to execute and the
instrument will be unresponsive during this period.
Before issuing additional commands, use the *OPC?
query command to determine if the operation is
complete (indicated by the return of the number "1").
The *OPC? query command can be appended to the
command as shown below.