Keysight Technologies N4960A User Manual page 146

Serial bert 17 and 32 gb/s
Table of Contents

Advertisement

Remote Operation
146
Command
:OUTSubrate:AMPLitude
Description
Set the amplitude of the divided clock output logic level from 0.3 V to
1.7 V in 0.005 V increments. The optional units are V (default) and mV.
The default value/unit is 0.7 V.
Example
:OUTS:AMPL 750mV
Command
:OUTSubrate:AMPLitude?
Description
Return the amplitude value of the divided clock output logic level. The
range is 0.3 V to 1.7 V.
Example
:OUTS:AMPL?
0.750V
Command
:OUTSubrate:COUPling
Description
Set the position of the internal switch to AC (default) or DC. Offset or
termination must be set to 0 V to change coupling to AC or a setting
conflict error message will be generated. If coupling is already set to AC,
it can be changed to DC without any dependencies.
Example
:OUTJ:COUP DC
Keysight N4960A Serial BERT 17 and 32 Gb/s User Guide