Keysight Technologies N4960A User Manual page 167

Serial bert 17 and 32 gb/s
Table of Contents

Advertisement

Remote Operation
Command
:PG:DATA:DEEMphasis:POST1?
Description
Return the value of the 5-tap data de-emphasis POST1 cursor. The de-
emphasis range is −30 dB to 0 dB. This command applies to the
N4951B-D17/-D32 only.
Example
:PG:DATA:DEEM:POST1? (@ 0)
−3.0dB
Queries the postcursor 1 setting for the pattern generator connected to
channel 0 (Jitter) only.
Command
:PG:DATA:DEEMphasis:POST2
Description
Set the 5-tap data de-emphasis POST2 cursor from −30 dB to +30 dB
(relative to post-cursor1) in 0.1 dB increments. The optional unit is dB.
This command applies to the N4951B-D17/-D32 only.
Example
:PG:DATA:DEEM:POST2 0
Sets the postcursor 2 to 0 dB for the pattern generator connected to
channel 0 (Jitter) only.
Keysight N4960A Serial BERT 17 and 32 Gb/s User Guide
167