Keysight Technologies N4960A User Manual page 132

Serial bert 17 and 32 gb/s
Table of Contents

Advertisement

Remote Operation
132
Command
:OUTJitter:JITTer:PERiodic
Description
Turn the periodic jitter output ON or OFF. The default is OFF.
Periodic jitter cannot be enabled if any of the other jitter sources are
enabled.
Example
:OUTJ:JITT:PER ON;*OPC?
1
Command
:OUTJitter:JITTer:PERiodic?
Description
Return the status of the periodic jitter. The returned string is either ON
or OFF.
Example
:OUTJ:JITT:PER?
ON
This command requires some time to execute and the
instrument will be unresponsive during this period.
Before issuing additional commands, use the *OPC?
query command to determine if the operation is
complete (indicated by the return of the number "1").
The *OPC? query command can be appended to the
command as shown below.
Keysight N4960A Serial BERT 17 and 32 Gb/s User Guide

Advertisement

Table of Contents
loading

Table of Contents