Keysight Technologies N4960A User Manual page 148

Serial bert 17 and 32 gb/s
Table of Contents

Advertisement

Remote Operation
148
Command
:OUTSubrate:OFFSet
Description
Set the offset voltage of the divided clock output logic level from –2.4 V
to +2.4 V in 0.005 V increments. The optional units are V (default) and
mV. The default value/unit is 0 V.
If the desired offset voltage is incompatible with the current termination
voltage, a setting conflict error message will be generated and the offset
cannot be set.
If the offset is changed to a value other than 0 V while the coupling is
set to AC, a setting conflict error message will be generated and the
offset cannot be set. If the coupling is set to DC, then offset can be
changed.
Example
:OUTS:OFFS 2V
Command
:OUTSubrate:OFFSet?
Description
Return the offset value of the divided clock output logic level. The range
is –2.4 V to +2.4 V.
Example
:OUTS:OFFS?
2.000V
Keysight N4960A Serial BERT 17 and 32 Gb/s User Guide