Remote Operation
126
Command
:OUTDelay:OFFSet
Description
Set the offset voltage of the output delay logic level from –2.4 V to +2.4
V in 0.005 V increments. The optional units are V (default) and mV. The
default value/unit is 0 V.
If the desired offset voltage is incompatible with the current termination
voltage, a setting conflict error message will be generated and the offset
cannot be set.
If the offset is changed to a value other than 0 V while the coupling is
set to AC, a setting conflict error message will be generated and the
offset cannot be set. If the coupling is set to DC, then offset can be
changed.
Example
:OUTD:OFFS 2V
Command
:OUTDelay:OFFSet?
Description
Return the offset value of the output delay logic level. The range is –2.4
V to +2.4 V.
Example
:OUTD:OFFS?
2.000V