Performance Specifications
Parameter
SJ
1
Frequency Range
SJ Modulation Range
RJ
1
Modulation Range
Low Frequency PJ
Modulation
1
Range
Low Frequency External
Modulation Range
High Frequency External
Modulation Range
Indicators
1
Jitter injection specifications (SJ, PJ, RJ, Ext) apply only to a pattern generator connected to the Jitter port.
110
Specification
1 Hz to 150 MHz, over-programmable to 200 MHz
Range of SJ1 and SJ2. The maximum combined peak jitter of SJ1 +
SJ2 +RJ (p-p) + external jitter are applied to the high frequency band
modulator (see
Figure
48).
0 to 0.8 UI p-p for modulation frequency 1 Hz to 30 MHz, 4 to
32 Gb/s
0 to 0.33 UI p-p for modulation frequency >30 MHz to 150 MHz, 4 to
29 Gb/s
0 to 0.2 UI p-p for modulation frequency >30 MHz to 150 MHz, >29 to
31.5 Gb/s
Over-programmable to 1.0 UI
0 to 24 mUI rms, 4 to 29 Gb/s
0 to 14 mUI rms, >29 to 31.5 Gb/s
Over-programmable to 50 mUI rms
0.002 to 100 UI for frequency ≤ 62.5 kHz
0.002 to (6.26E6 / PJ frequency) for frequency > 62.5 kHz to 17 MHz
Figure
48.
See
0.001 to 50 UI for frequency ≤ 68.4 kHz
0.001 to (3.42E6 / modulation frequency) for frequency > 68.4 kHz to
4 MHz
See
Figure
48.
See
Figure
48.
Ch ID – connected to N4960A channel
Atten – error condition occurred and logged in Error Log
On – data output on