Keysight Technologies N4960A User Manual page 185

Serial bert 17 and 32 gb/s
Table of Contents

Advertisement

Remote Operation
Command
:ED:DATA:AALign:CAALign:DSTep?
Description
Return the sampling delay step size. The delay step size is fixed at
0.02 UI.
Example
:ED:DATA:AAL:CAAL:DST?
0.002UI
Command
:ED:DATA:AALign:CAALign:SVOLtage
Description
Turn the sampling threshold voltage ON or OFF. This determines
whether the auto alignment finds the optimal sampling threshold
voltage (eye height). The default is ON. The sampling threshold voltage
result will not be available if SVOL is turned OFF.
Example
:ED:DATA:AAL:CAAL:SVOL on
Command
:ED:DATA:AALign:CAALign:SVOLtage?
Description
Return the status of the sampling threshold voltage. The returned string
is either ON or OFF.
Example
:ED:DATA:AAL:CAAL:SVOL?
ON