Remote Operation
150
Command
:OUTSubrate:TERMination
Description
Set the termination voltage of the divided clock output logic level from –
2.4 V to +2.4 V in 0.005 V increments. The optional units are V (default)
and mV. The default value/unit is 0 V.
If the desired termination voltage is incompatible with the current offset
voltage, a setting conflict error message will be generated and the
termination cannot be set.
If the termination is changed to a value other than 0 V while the
coupling is set to AC, a setting conflict error message will be generated
and the termination cannot be set. If the coupling is set to DC, then
termination can be changed.
Example
:OUTS:TERM 1.3V
Command
:OUTSubrate:TERMination?
Description
Return the termination value of the subrate output logic level. The range
is –2.4 V to +2.4 V.
Example
:OUTS:TERM?
1.300V
Keysight N4960A Serial BERT 17 and 32 Gb/s User Guide