The O&P professional's setting is located in the middle of the scale. After making adjustments, this setting can
be restored by tapping the "Standard" button in the Cockpit app.
Prosthesis settings should be optimised using the adjustment software. The Cockpit app is not intended for
use by the O&P professional to set up the prosthesis. The patient can use the app to change the behaviour of
the prosthesis to a certain extent during everyday use (e.g. while becoming accustomed to the prosthesis). The
O&P professional can use the adjustment software to trace these changes at the patient's next appointment.
If the settings of a MyMode are to be modified, one must first switch to this MyMode.

9.4.1 Changing the prosthesis setting using the cockpit app

6
Product
MyModes
Functions
Heel Height
Settings
Status
General
Manage components
Imprint/Info

9.4.2 Overview of adjustment parameters in basic mode

The parameters in basic mode describe the dynamic behaviour of the prosthesis in a normal gait cycle. These para­
meters act as basic settings for automatically adjusting the damping behaviour to the current motion situation (e.g.
ramps, slow walking speed, etc.).
The following parameters can be modified:
Parameter
Pitch
Volume
Heel Resist.
Rollover Resist.
Stair Function
1B1-2=* Meridium
1) Once the component is connected and in the desired mode, tap the
 symbol in the main menu.
→ The navigation menu opens.
2) Tap the "Settings" menu option.
change
→ A list appears with the parameters for the currently selected mode.
3) Change the setting of the desired parameter by tapping the "<", ">"
2. Golf
symbols.
INFORMATION: The O&P professional's setting is marked and,
after the setting has been changed, can be restored by tapping
the "Standard" button.
Adjustment soft­
ware range
1000 Hz–4000 Hz 1000 Hz–4000 Hz Pitch (frequency) of beep signal for confirma­
0–4
10–60
110–170
ON–OFF
Setting range,
Meaning
app
tion tones
0–4
Volume of beep signal for confirmation tones
(e.g. when checking the charge level, switch­
ing MyModes).
The "0" setting deactivates the audible feed­
back signals. However, warning signals are
still generated when errors occur (see
Page 34).
± 20
Plantar flexion damping.
How fast the forefoot is lowered when weight
is placed on the heel.
± 10
This parameter defines how easy the rollover
is.
ON–OFF
Switching on this function increases the
rollover angle when walking down stairs. In
Use