DR6500
• Blocking Distance (C1.5)
• Reference Offset (C1.10)
• Display readings in Normal mode:
Display readings in Normal mode: 1st Meas. Page (C6.4) menu
Display readings in Normal mode:
Display readings in Normal mode:
• Display readings in Normal mode:
Display readings in Normal mode: 2nd Meas. Page (C6.5) menu
Display readings in Normal mode:
Display readings in Normal mode:
The flange facing is the reference point for distance measurement (0 m / 0 ft / 0¨). The position of
the measurement scale (specified by the 0% Range and 100% Range settings) is related to this
reference point. If you configured a measurement scale for the current output signal in the C4.1
Current output 1 menu, you can use a "standard scale" or a "reversed scale". On the standard
scale, the 0% Range measurement value agrees with an output of 4 mA and the 100% Range
measurement value agrees with an output of 20 mA. On the reversed scale, the 0% Range
measurement value agrees with an output of 20 mA and the 100% Range measurement value
agrees with an output of 4 mA.
You can change the reference point from which distance is measured. Use this menu
item:
• Reference Offset (C1.10)
INFORMATION!
Menu item C1.10 Reference Offset
Menu item C1.10 Reference Offset
Menu item C1.10 Reference Offset
Menu item C1.10 Reference Offset
If you move the reference point above the flange, make sure that you add the value given in menu
item C1.10 Reference Offset when you give a distance for the C4.1.2 0% Range and C4.1.3 100%
Range settings. If you move the reference point below the flange, subtract the value given in
menu item C1.10 Reference Offset when you give a distance for the C4.1.2 0% Range and C4.1.3
100% Range settings.
CAUTION!
If C4.1.1 Current Out. 1 Var. is set to "Distance" and C4.1.2 0% Range (standard scale) is set in
the blocking distance, then the device will not be able to use the full current output range.
DR6500-PM-English-Issue 1