DR6500
Description of errors and corrective actions
Status
Error message
type
F F F F
Sensor
Sensor
Sensor
Sensor
Corrupt Sensor Parameter
No Signal
Microwave Tuning Voltage Error
Sensor Comm. Error
No measurement value
F F F F
Electronics
Electronics
Electronics
Electronics
Fatal Converter Error (DM)
Fatal Converter Error (CO)
Fatal Converter Error (Generic)
CO Safety Reaction
Internal Comm. Error
Power Supply Error
Sensor/Conv. FW Mismatch
Max. Number of Restarts
Inconsistent Parameters
Sensor Comm. Error
DR6500-PM-English-Issue 1
Description
The sensor memory is bad.
There is no signal from the
antenna or the antenna signal is
too weak for a long time.
A microwave error occurred.
Internal bus communication error
or hardware failure.
The converter receives no
measurement data for a long time.
An electronics or hardware failure
1
occurred.
An electronics or hardware failure
2
occurred.
An electronics or hardware failure
occurred.
Current Output Safety Reaction.
3
This error message is shown if the
current output is less than 3.6 mA
or more than 21 mA.
Internal bus communication error. De-energize and then energize the
Internal voltage is too low to
energize the sensor.
Internal voltage is too high.
The sensor firmware version does
not agree with the converter
firmware version. This can be
caused by a bad firmware update.
The device could not start the
measuring mode after a given
number of times that the device
was de-energized and then
energized again.
Internal bus communication error. De-energize and then energize the
Internal bus communication error
or hardware failure.
www.drexelbrook.com
OPERATION
Corrective action
De-energize and then energize the
device again.
If the message is shown again, tell
the supplier.
Do a check of the device
installation. For more data about
installation, refer to the handbook.
If the message is shown again, tell
the supplier.
De-energize and then energize the
device again.
If the message is shown again, tell
the supplier.
Do a check of the device
installation. For more data about
installation, refer to the handbook.
If the message is shown again, tell
the supplier.
De-energize and then energize the
device again.
If the message is shown again, tell
the supplier.
Do a check for other error
messages.
device again.
If the message is shown again, tell
the supplier.
Check the electrical connections
to the power supply connection or
replace the converter.
De-energize and then energize the
device again.
If the message is shown again, tell
the supplier.
Do a check of the power supply.
device again.
If the message is shown again, tell
the supplier.
6
107