Ametek DR6500 Handbook page 88

80 ghz radar (fmcw) level transmitter for powders and dusty atmosphere
Table of Contents

Advertisement

OPERATION
6
Menu
Function
No.
C6.5.8
3rd Value
Variable
C6.5.9
Format 3rd
Value
C7 Device
C7.1 Information
C7.1 Information
C7.1 Information
C7.1 Information
C7.1.1
Tag
C7.1.2
Serial Number
C7.1.3
Device Name
C7.1.4
V Number
C7.1.5
Electronic
Revision
C7.1.6
Software
Revision
C7.1.7
Electronic
Serial No.
C7.1.8
Production Date This the date that the manufacturer finished
C7.1.9
Calibration Date This the date that the manufacturer
C7.2 Security
C7.2 Security
C7.2 Security
C7.2 Security
C7.2.1
Login
C7.2.2
Change
Password
88
Function description
This changes the measurement type of the
third value on the measurement page. This
menu item is only available if you set C6.5.1
Function to "Three Values".
Minimum access level to change the
Minimum access level to change the
Minimum access level to change the
Minimum access level to change the
setting: User
setting:
setting:
setting:
You can change the number of decimal
places that are shown for the second value
on the measurement page in Normal mode.
For example, if you set this menu item to
"X.XX", the value shown will be corrected to
2 decimal places. If you set this menu item
to "Automatic", the device will adjust the
number of decimal places automatically.
Minimum access level to change the
Minimum access level to change the
Minimum access level to change the
Minimum access level to change the
setting: User
setting:
setting:
setting:
Use this menu item to read the Tag name.
This is the device fabrication number
This gives the device family name and model
code.
This is the manufacturer's option code for
device configuration.
This is the hardware revision number. This
number agrees with NAMUR NE 53
Guidelines.
This is the software revision number. This
number agrees with NAMUR NE 53
Guidelines.
This is the hardware fabrication number.
This number agrees with NAMUR NE 53
Guidelines.
the device. The date format is: Year-Month-
Day.
calibrated the device. The date format is:
Year-Month-Day.
Enter the appropriate password here to
change settings. If you do not enter the
password, you can only change settings for
the "user" access level. For more data and
the default passwords for the "operator"
and "expert" access levels, refer to
Protection of the device settings (access
levels)
on page 59.
This changes the password for the
"operator" and "expert" access levels. For
the procedure to change the password,
Protection of the device settings
refer to
(access levels)
on page 59 (How to change
the password).
www.drexelbrook.com
Selection list
Level, Distance, Sensor
Value, Reflection
2
X, X.X, X.XX, X.XXX,
X.XXXX, X.XXXXX,
X.XXXXXXX, Automatic
Read only
Read only
Read only
Read only
Read only
Read only
Read only
Read only
Read only
4-digit hexadecimal
password
4-digit hexadecimal
password
DR6500-PM-English-Issue 1
DR6500
Default
Reflection
Automatic
TANK01
1
Refer to
"Function
description"
Refer to
"Function
description"

Advertisement