Measurement data format errors
Figure 6-2: Error symbol: the number of digits and decimal places is not sufficient for the measurement data
1 Error symbol: the number of digits and decimal places is not sufficient for the measurement data. It is possible that
the length units must be changed from "mm" to "m".
In this example, the device measures a distance of 10.001 m, but C7.5.1 Length is set to "mm"
and C6.4.5 Format 1st Value is set to "X.XXX" (four digits with three decimal places). This is not
sufficient to show a measurement value of 10.001 m. If the measurement value is equal to or
more than 10 m, set C6.4.5 Format 1st Value to "Automatic".
You can change the number of digits and decimal places in the measurement values shown in
Normal mode.
INFORMATION!
How to change the number of digits and decimal places in the measurement values shown in
How to change the number of digits and decimal places in the measurement values shown in
How to change the number of digits and decimal places in the measurement values shown in
How to change the number of digits and decimal places in the measurement values shown in
Normal mode
Normal mode
Normal mode
Normal mode
1st Measurement Page:
1st Measurement Page:
1st Measurement Page: Go to menu C6.4 1st Measur. Page
1st Measurement Page:
and decimal places in C6.4.5 Format 1st value
value
value.
value
value
2nd Measurement Page: Go to menu C6.5 2nd Measur. Page
2nd Measurement Page:
2nd Measurement Page:
2nd Measurement Page:
digits and decimal places in C6.5.5 Format 1st value
3rd value
3rd value
3rd value.
3rd value
If there is a large change in a measurement value, set the related menu item to "Automatic".
Volume or mass measurement
You must make a conversion table (strapping table) to show measurement data as a volume or a
mass. Go to C3.2 Input table (Full Setup > Conversion
data, refer to
Functions of keypad buttons (Normal mode)
Keypad button
56
Format 1st value
Format 1st value
Format 1st value, C6.4.7 Format 2nd value
Full Setup > Conversion) to make the strapping table. For more
Full Setup > Conversion
Full Setup > Conversion
How to configure the device to measure volume or mass
Symbol
[> > > > ]
[Right]
[^ ^ ^ ^ ]
[Return]
[> > > > ]+[ ]
[Escape]
[ ]
[Down]
[ ]
[Up]
1st Measur. Page and then change the number of digits
1st Measur. Page
1st Measur. Page
Format 2nd value
Format 2nd value or C6.4.9 Format 3rd
Format 2nd value
2nd Measur. Page and then change the number of
2nd Measur. Page
2nd Measur. Page
Format 1st value, C6.5.7 Format 2nd value
Format 1st value
Format 1st value
Enter Program mode
Change screen (measurement pages 1 and 2, and the status
message page)
Change screen (measurement pages 1 and 2 and the status
message page)
www.drexelbrook.com
Format 3rd
Format 3rd
Format 3rd
Format 2nd value or C6.5.9 Format
Format 2nd value
Format 2nd value
on page 101.
Function
DR6500-PM-English-Issue 1
DR6500
Format
Format
Format

Advertisement

Table of Contents
loading

Table of Contents