How To Tilt The Device; How To Attach The Purging System Adaptor To The Device - Ametek DR6500 Handbook

80 ghz radar (fmcw) level transmitter for powders and dusty atmosphere
Table of Contents

Advertisement

DR6500
• Put the device carefully on the orientation system. Turn the threaded connection on the
antenna to attach the device to the orientation system. Tighten the connection to the correct
torque (not more than 40 N·m / 29.5 lbf·ft) with a 50 mm (G 1½ / 1½ NPT) or 65 mm (G 3 /
3 NPT) open-end wrench.
i
End of the procedure.
INFORMATION!
If it is necessary to send an order for the orientation system, refer to Accessories: process
connection kits on page 158

3.8.2 How to tilt the device

If you must measure a product which does not have a level surface then attach an orientation
system and tilt the device to get a good measurement signal. You can tilt the device to a
maximum angle of 30° from the vertical axis.
Figure 3-13: How to tilt the device
Equipment needed:
• 5 mm Allen wrench (not supplied)
• Loosen the 4 socket head screws around the orientation system with a 5 mm Allen wrench.
• Tilt the device to the correct angle.
• Tighten the 4 socket head screws. This step locks the orientation system to prevent
movement.

3.9 How to attach the purging system adaptor to the device

If it is necessary to clean the Lens antenna, you can attach a purging system adaptor to the
device. This part is available as an accessory for devices with threaded connections.
INFORMATION!
The purging system adaptor has an inlet with a G 1/4 threaded connection.
DR6500-PM-English-Issue 1
.
www.drexelbrook.com
INSTALLATION
3
33