Making Time-Gated Measurements

Gated FFT Measurement

Gated FFT Measurement
This procedure utilizes gated FFT to gate the FM signal. For concept and theory
information about gated FFT see
Step
1 Set the analyzer to the
Spectrum Analyzer mode.
2 Preset the analyzer
3 Set the center frequency,
span, reference level and
attenuation.
4 Set the Gate source to the
external trigger input on the
rear panel.
5 Set the gate method and turn
gate on.
6 Select the minimum
resolution bandwidth
required.
"How time gating works" on page
Action
Press Mode, Spectrum
Analyzer.
Press Mode Preset.
a. Press FREQ Channel,
Center Freq, 40, MHz.
b. Press SPAN X Scale,
Span, 500, kHz.
c. Press AMPTD Y Scale,
Ref Level, 10, dBm.
Press Sweep/Control,
Gate, More, Gate
Source, External 1.
a. Press Sweep/Control,
Gate, Gate Method,
FFT.
b. Press Gate (On).
Press BW, Res BW
(Auto).
108
Notes
This enables the spectrum analyzer
measurements.
See
Figure
8-12.
199.