Getting Started with the Spectrum Analyzer Measurement Application
Making a Basic Measurement
Viewing a signal
Step
4 Return the current
5 Route the internal
6 Set the reference
7 Set the center
8 Set the frequency
Reading frequency and amplitude
Step
1 Activate a marker
2 To return the marker
Changing reference level
Step
1 Change the reference
mode settings to
factory defaults.
50 MHz signal to the
analyzer input.
level to 10 dBm.
frequency to 40
MHz.
span to 50 MHz.
and place it on the
highest amplitude
signal.
to the peak of the
signal.
level.
Action
•
Press Mode Preset.
•
Press Input/Output,
RF Calibrator, 50,
MHz.
•
Press AMPTD Y
Scale, 10, dBm.
•
Press FREQ
Channel, Center
Freq, 40, MHz.
•
Press SPAN, 50, MHz.
Action
•
Press Peak
Search.
•
Press Peak
Search.
Action
a. Press AMPLTD
Y Scale.
b. Press Marker −>,
Mkr −> Ref Lvl.
14
Notes
The 50 MHz reference
signal appears on the
display
Notes
The frequency and amplitude of
the marker appear in the active
function block in the
upper-right of the display. You
can use the knob, the arrow
keys, or the softkeys in the Peak
Search menu to move the
marker around on the signal.
Notes
The reference level is now the
active function.