Demodulating AM Signals

Measuring the Modulation Index of an AM Signal

Measuring the Modulation Index of an AM Signal
This procedure demonstrates how to use the signal analyzer as a fixed-tuned
(time-domain) receiver to measure the modulation index as a percent AM value of an
AM signal.
Step
1 Set up the signal source.
2 Connect an Keysight
ESG RF signal source to
the analyzer RF INPUT
as shown.
3 Select the mode.
4 Preset the analyzer.
5 Set the center frequency,
span, RBW and the
sweep time.
Action
a. Set the source frequency to
300 MHz.
b. Set the source amplitudes to
−10 dBm.
c. Set the AM depth to 80%.
d. Set the AM rate to 1 kHz.
e. Turn AM on.
Press Mode, Spectrum
Analyzer.
Press Mode Preset.
a. Press FREQ Channel,
Center Freq, 300, MHz.
b. Press SPAN X Scale, Span,
500, kHz.
c. Press BW, Res BW, 30,
kHz.
d. Press Sweep/Control,
Sweep Time, 20, ms.