Equipment Set Up - Keysight X Series Measurement Manual

Signal analyzer spectrum analyzer mode
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Avoiding waveguide flange damage
Install the waveguide flange cap whenever the mixer is not connected to a device
under test. This will protect the waveguide flange mating surface.
Mixer waveguide connections
Assure the shoulder of the mixer waveguide flange is properly aligned with the
flange of the device under test. To ensure proper mating, it is important to tighten all
screws equally. To do this, tighten opposed screws in pairs by a small amount until
all are snug. Final torque must not exceed 7 in pounds.
For additional information regarding use with a particular X-Series analyzer, refer to
that analyzer's User's and Programmer's Reference Guide.

Equipment set up

The following examples explain how to connect the external mixers to the signal
analyzer, and how to use the signal-identification functions.
Step
Action
a. Connect an SMA
1 SMA connection
a. Connect a USB
2 USB connection
Using the M1970 Series Mixers with X-Series Signal Analyzers (Option EXM)
Notes
The torque for the SMA cables and adapters should not
cable from the mixer
exceed 8 in-lbs.
to the spectrum
The LO/IF connection between the mixer and the signal
analyzer EXT MIXER
analyzer must be in place before the USB cable from the
front panel SMA.
mixer is connected. Connecting the USB cable
automatically triggers the LO adjustment, and if the LO/IF
cable is not connected the adjustment will not complete
and an error will occur.
When a connection is made, the green LED on the mixer
cable from the mixer
lights up indicating that the mixer has power and the
to the spectrum
processor inside the mixer is running. The spectrum
analyzer.
analyzer is automatically switched to external mixing
mode, and the mixer model number and mixer option are
used to set the start and stop frequencies for the mixing
band.
Option Esc External Source Control