Comparing Signals Not On The Same Screen Using Marker Delta - Keysight X Series Measurement Manual

Signal analyzer spectrum analyzer mode
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Comparing Signals not on the Same Screen Using Marker Delta

Measure the frequency and amplitude difference between two signals that do not
appear on the screen at one time. (This technique is useful for harmonic distortion
tests when narrow span and narrow bandwidth are necessary to measure the low
level harmonics.)
In this procedure, the analyzer 10 MHz signal is used to measure frequency and
amplitude differences between one signal on screen and one signal off screen. Delta
marker is used to demonstrate this comparison.
Figure 2-3
Comparing One Signal on Screen with One Signal Off Screen
Step
1 Connect the 10 MHz OUT
from the rear panel to the
front panel RF input.
2 Select the mode.
3 Preset the mode.
4 Configure the analyzer
settings.
Comparing Signals not on the Same Screen Using Marker Delta
Action
Press Mode, Spectrum
Analyzer.
Press Mode Preset.
a. Press FREQ Channel,
Center Freq, 10, MHz.
b. Press SPAN X Scale, Span,
5, MHz.
c. Press AMPTD Y Scale, Ref
Level, 10, dBm.
25
Measuring Multiple Signals
Notes
This sets the analyzer center frequency,
span and reference level to view the 10
MHz signal and its harmonics up to 50
MHz.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

N9030aN9010aN9038aN9000aN9020a