Making Time-Gated Measurements

Generating a Pulsed-RF FM Signal

Generating a Pulsed-RF FM Signal
When performing these measurements you can use a digitizing oscillascope or your
Keysight X-Series Signal Analyzer (using Gate View) to set up the gated signal.
Refer back to these first three steps to set up the pulse signal, the pulsed-RF FM
signal and the oscilloscope settings when performing the gated LO procedure
(page
(page
For an instrument block diagram and instrument connections see
Instruments to Make Time-Gated Measurements" on page

Signal source setup

Step 1. Set up the pulse signal with a period of 5 ms and a width of 4 ms:
There are many ways to create a pulse signal. This example demonstrates how to
create a pulse signal using a pulse generator or by using the internal function
generator in the ESG.See
you are using a second ESG. Select either the pulse generator or a second ESG to
create the pulse signal.
Table 8-1
81100 Family Pulse Generator Settings
Period
Pulse width
High output level
Waveform
Low output level
Delay
Table 8-2
ESG #2 Internal Function Generator (LF OUT) Settings
LF Out Source
LF Out Waveform
LF Out Period
LF Out Width (pulse
width)
LF Out Amplitude
LF Out
RF On/Off
Mod On/Off
100), the gated video procedure
108).
Table 8-1
5 ms (or pulse frequency equal to 200 Hz)
4 ms
2.5 V
pulse
-2.5 V
0 or minimum
FuncGen
Pulse
5 ms
4 ms
2.5 Vp
On
Off
On
94
104) and gated FFT procedure
(page
if you are using a pulse generator or
"Connecting the
99.
Table 8-2