Measuring Digital Communications Signals

Channel Power Measurements

Channel Power Measurements
This section explains how to make a channel power measurement on a W-CDMA
(3GPP) mobile station. (A signal generator is used to simulate a base station.) This
test measures the total RF power present in the channel. The results are displayed
graphically as well as in total power (dB) and power spectral density (dBm/Hz).
Step
1 Set up the signal
generator.
2 Connect the source RF
OUTPUT to the
analyzer RF INPUT as
shown.
3 Select the mode.
4 Preset the analyzer.
5 Set the radio standard
and toggle the device to
mobile station.
6 Set the center frequency.
7 Initiate the channel
power measurement.
Action
a. Set the mode to W-CDMA.
b. Set the frequency to 1.920
GHz.
c. Set the amplitude to
−20 dBm.
Press Mode, Spectrum
Analyzer.
Press Mode Preset.
Press Mode Setup, Radio
Std, 3GPP W-CDMA,
3GPP W-CDMA, Device
(MS).
Press FREQ Channel, 1.920,
GHz.
Press Meas, Channel Power.
112
Notes
The Channel Power measurement result
should look like
"Channel Power
Measurement Result." on
page
113.