Option Esc External Source Control
Using Option ESC with the Keysight MXG Signal Sources.
Using Option ESC with the Keysight MXG Signal Sources.
The following examples explain how to connect the MXG signal source to the signal
analyzer.
Step
1 Set up the equipment
MXG
2 Set the analyzer to the
Spectrum Analyzer mode
3 Preset the analyzer
4 Select external mixing.
5 To select Q, U, V or W
bands
6 View the spectrum display
7 Tune the analyzer to the
input signal frequency
Action
•
Connect the signal source to the
signal analyzer as shown:
•
Press Mode, Spectrum Analyzer
•
Press Mode Preset.
•
Press Input/Output, External
Mixer, External Mixer, External
Mixer Setup
a. Press Mixer Presets, Keysight
11970.
b. Select the appropriate key
corresponding to 11970 frequency
band.
•
Press Return twice.
a. Press FREQ Channel.
b. Enter a center frequency or a start
and stop frequency
176
Notes
The X-Series controls the MXG
via the virtual instrument
software architecture interface
(VISA), which uses a connection
of
"Lan" on page
177, GPIB,
or USB.
X-Series
The default is 11970A