Step
8 Turn on the Signal ID
function

Lan

The X-Series with Option ESC controls the MXG through a LAN connection based
on TCP/IP protocol. The TCP/IP protocol can only be established with correct IP
addressing.
NOTE
You must be logged in as an Administrator to change the TCP/IP properties. This
requires the use of a USB mouse and keyboard.
X-Series Instrument
Step
1 Log off as default user.
2 Log on as an administrator.
3 Assign an IP adress and a subnet mask.
NOTE
Assure only ONE correction file is turned on because it is possible to turn on
multiple correction files, and if some of the files share the same frequency points, the
correction that results for those shared frequency points will cause measurement
errors. Therefore assure that only the correction file required for the measurement is
turned on, and turn off all other corrections.

Loading conversion loss data for the PXA Signal Analyzer

The conversion loss data supplied with your mixer can be loaded into your signal
analyzer from one of two sources:
Using Option ESC with the Keysight MXG Signal Sources.
Action
Press Input Output, External
Mixing, Signal ID Mode, Image
Suppress, Signal ID On.
177
Option Esc External Source Control
Notes
This enables you to identify true
signals from images and
harmonics.
See
"Signal ID" on page 182
for more information.
Action
Click Start, Log Off, Log Off
At the log in prompt enter:
User name: administrator