Step
7 Set the RBW to auto,
gate view to off, gate
method to LO, and
gate to on.
Figure 8-2
Gated RF Signal with Auto RBW
Digitizing oscilloscope setup
If you are using a digitizing oscillascope, set up the oscilloscope to view the trigger,
gate and RF signals (see
Table 8-4
Keysight Infiniium Oscilloscope with 3 or more input channels: Instrument
Connections
Timebase
Channel 1
Action
a. Press Sweep/Control, Gate, Gate
View (Off).
b. Press BW, Res BW (Auto).
c. Press Sweep/Control, Gate, Gate
Method, LO.
d. Press Gate (On).
Figure 8-3
1 ms/div
ON, 2 V/div, OFFSET = 2 V, DC coupled, 1 M Ω input,
connect to the pulse signal (ESG LF OUTPUT or pulse
generator OUTPUT). Adjust channel 1 settings as
necessary.
Making Time-Gated Measurements
Generating a Pulsed-RF FM Signal
Notes
for an example of the oscilloscope display):
97