Occupied Bandwidth Measurements - Keysight X Series Measurement Manual

Signal analyzer spectrum analyzer mode
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Measuring Digital Communications Signals

Occupied Bandwidth Measurements

Occupied Bandwidth Measurements
This section explains how to make the occupied bandwidth measurement on a
W-CDMA (3GPP) mobile station. (A signal generator is used to simulate a base
station.) The instrument measures power across the band, and then calculates its
99.0% power bandwidth.
Step
1 Set up the signal generator.
2 Connect the source RF
OUTPUT to the analyzer
RF INPUT as shown.
3 Select the mode.
4 Preset the analyzer.
5 Set the radio standard and
toggle the device to mobile
station.
6 Set the center frequency.
7 Initiate the occupied
bandwidth measurement.
Action
a. Set the mode to
W-CDMA.
b. Set the frequency to 1.920
GHz.
c. Set the amplitude to
−20 dBm.
Press Mode, Spectrum
Analyzer.
Press Mode Preset.
Press Mode Setup, Radio
Std, 3GPP W-CDMA,
3GPP W-CDMA, Device
(MS).
Press FREQ Channel,
1.920, GHz.
Press Meas, Occupied
BW.
114
Notes
The Occupied BW measurement result
should look like
"Occupied BW
Measurement Result" on page
115.

Hide quick links: