Radiation and General Safety
Radiation Profile
Radiation Profile
Table 3-4. Primary Beam Dose Rates in mSv/hr
Max Power
Max Power
Settings
Settings
kVp
µA
40
50
40
50
50
40
20
100
15
100
Table 3-5. Primary Beam Dose Rates in Rem/hr
Max Power
Max Power
Settings
Settings
kVp
µA
40
50
40
50
50
40
20
100
15
100
3-10
NITON XL3 Analyzer User's Guide
Table
3-4,
Table
radiation dose rates that are present at various points in space around the
NITON XL3t analyzer when it is being used.
these dose rate points are relative to the analyzer.
Window
Contact
Deep
Max in Following
(mSv/hr)
Modes* (Filter)
G, D, E, A, P, F, M, J
1,410
(Main Filter)
B (Main Filter)
750
H (Main Filter), D, E,
1,090
M, J (High Filter)
D, H, E, M, J (Low
1,450
Filter)
A, B (Low Filter)
133
Window
Contact
Deep
Max in Following
(Rem/hr)
Modes* (Filter)
G, D, E, A, P, F, M, J
141
(Main Filter)
B (Main Filter)
75
H (Main Filter), D, E,
109
M, J (High Filter)
D, H, E, M, J (Low
145
Filter)
A, B (Low Filter)
13.3
* G = Alloy, B = Alloy Electronics, F = Dental Alloy, P = Precious Metals, M = Mining, D = Soil,
J = Exploration, A = Lead Paint, E = Thin Sample, H = Plastic
3-5,
Table 3-6
and
Table 3-7
Window
Contact
Shallow
(mSv/hr)
1,410
2,250
4,060
31,717
10,567
Window
Contact
Shallow
(Rem/hr)
141
225
406
3,171.7
1,056.7
below describes the external
Figure 3-4
illustrates where
5 cm
30 cm
Deep
Deep
(mSv/hr)
(mSv/hr)
50.00
6.3
40.00
5.4
84.0
12.2
5.2
0.5
4.3
0.42
5 cm
30 cm
Deep
Deep
(Rem/hr)
(Rem/hr)
5.00
0.63
4.00
0.54
8.4
1.22
0.52
0.05
0.43
0.042
Thermo Scientific