Configure Density Measurement; Configure Density Measurement Unit - Emerson Micro Motion 1500 Configuration And Use Manual

Transmitters with analog outputs
Hide thumbs Also See for Micro Motion 1500:
Table of Contents

Advertisement

Configure process measurement
Table 4-3:
Flow Direction setting
Forward
Reverse
Bidirectional
Absolute Value
Negate Forward
Negate Bidirectional
(1) Refer to the digital communications status bits for an indication of whether flow is positive or negative.
Effect of flow direction on flow totals
Flow direction affects how flow totals and inventories are calculated.
Flow Direction setting
Forward
Reverse
Bidirectional
Absolute Value
Negate Forward
Negate Bidirectional
4.5

Configure density measurement

The density measurement parameters control how density is measured and reported.
4.5.1

Configure Density Measurement Unit

ProLink III
Field Communicator Configure > Manual Setup > Measurements > Density > Density Unit
Overview
Density Measurement Unit controls the measurement units that will be used in density
calculations and reporting.
38
Effect of the flow direction on flow values
Forward
Totals increase
Totals do not change
Totals increase
Totals increase
Totals do not change
Totals decrease
Device Tools > Configuration > Process Measurement > Density
Micro Motion Model 1500 Transmitters with Analog Outputs
Actual flow direction
Forward
Zero flow
Positive
0
Positive
0
Positive
0
(1)
Positive
0
Negative
0
Negative
0
Actual flow direction
Zero flow
Totals do not change
Totals do not change
Totals do not change
Totals do not change
Totals do not change
Totals do not change
Reverse
Negative
Negative
Negative
(1)
Positive
Positive
Positive
Reverse
Totals do not change
Totals increase
Totals decrease
Totals increase
Totals increase
Totals increase

Advertisement

Table of Contents
loading

Table of Contents