9.2.5

Schedule automatic execution of the SMV test

You can set up and run a single test at a user-defined future time. You can also set up and
run tests on a regular schedule.
Manage scheduled test execution using ProLink III
1.
2.
3.
4.
Manage scheduled test execution using the
Field Communicator
1.
2.
3.
4.
5.
9.3

Use PVR, TBR, and TMR

For detailed information about Production Volume Reconciliation (PVR), Transient Mist
Remediation (TMR), and Transient Bubble Remediation (TBR), see the Micro Motion
and Gas Production Applications manual.
Restriction
PVR, TBR, and TMR are available only with a transmitter that is ordered with one of these software
options:
Configuration and Use Manual
Choose Device Tools > Diagnostics > Meter Verification > Schedule Meter
Verification.
To schedule a single test or the first test in recurring execution, specify a value for
Hours Until Next Run.
To schedule recurring execution, specify a value for Hours Between Recurring Runs.
To disable scheduled execution:
To disable execution of a single scheduled test, set Hours Until Next Run to 0.
To disable recurring execution, set Hours Between Recurring Runs to 0.
To disable all scheduled execution, click Disable Scheduled Execution.
Navigate to the Smart Meter Verification menu:
Overview > Shortcuts > Meter Verification
Service Tools > Maintenance > Routine Maintenance > Meter Verification
Choose Automatic Verification.
To schedule a single test or the first test in recurring execution, specify a value for
Hrs Until Next Run.
To schedule recurring execution, specify a value for Set Recurring Hours.
To disable scheduled execution:
To disable execution of a single scheduled test, set Hrs Until Next Run to 0.
To disable recurring execution, set Set Recurring Hours to 0.
To disable all scheduled execution, choose Turn Off Schedule.
Measurement support
®
Oil
109