Emerson Micro Motion 1500 Configuration And Use Manual page 179

Transmitters with analog outputs
Hide thumbs Also See for Micro Motion 1500:
Table of Contents

Advertisement

Table C-1:
Transmitter default values and ranges
Type
Parameter
Base mass time
Mass flow conversion factor
Base volume unit
Base volume time
Volume flow conversion factor
Variable map-
Primary variable
ping
Secondary variable
Tertiary variable
Quaternary variable
mA Output 1
Primary variable
LRV
URV
AO cutoff
AO added damping
LSL
USL
MinSpan
Fault action
AO fault level – downscale
AO fault level – upscale
Last measured value timeout
mA Output 2
Secondary variable
LRV
URV
AO cutoff
AO added damping
Configuration and Use Manual
(continued)
Default
Range
sec
1
L
sec
1
Mass flow
Volume flow
Mass flow
Volume flow
Mass flow
–200.00000
g/s
200.00000 g/s
0.00000 g/s
0.00000 sec
–200 g/s
200 g/s
0.3 g/s
Downscale
2.0 mA
1.0 – 3.6 mA
22 mA
21.0 – 24.0 mA
0.00 sec
Density
0.00 g/cm3
10.00 g/cm3
Not-A-Number
0.00000 sec
Default values and ranges
Comments
The user-entered value is correc-
ted down to the nearest lower
value in a list of preset values.
Read-only.
LSL is calculated based on the
sensor size and characterization
parameters.
Read only.
USL is calculated based on the
sensor size and characterization
parameters.
Read-only.
173