Configure Flow Damping

ProLink III
Field Communicator Configure > Manual Setup > Measurements > Flow > Flow Damping
Overview
Damping is used to smooth out small, rapid fluctuations in process measurement.
Damping Value specifies the time period (in seconds) over which the transmitter will
spread changes in the process variable. At the end of the interval, the internal value will
reflect 63% of the change in the actual measured value.
Procedure
Set Flow Damping to the value you want to use.
Configuration and Use Manual
Specify Base Time Unit.
Base Time Unit is the existing time unit that the special unit will be based on.
Calculate Mass Flow Conversion Factor as follows:
a. x base units = y special units
b. Mass Flow Conversion Factor = x ÷ y
The original mass flow rate value is divided by this value.
Enter Mass Flow Conversion Factor.
Set Mass Flow Label to the name you want to use for the mass flow unit.
Set Mass Total Label to the name you want to use for the mass total and mass
inventory unit.
Set Base Mass Unit to Pounds (lb).
Set Base Time Unit to Seconds (sec).
Calculate Mass Flow Conversion Factor:
a. 1 lb/sec = 16 oz/sec
b. Mass Flow Conversion Factor = 1 ÷ 16 = 0.0625
Set Mass Flow Conversion Factor to 0.0625.
Set Mass Flow Label to oz/sec.
Set Mass Total Label to oz.