Emerson Micro Motion 1500 Configuration And Use Manual page 61

Transmitters with analog outputs
Hide thumbs Also See for Micro Motion 1500:
Table of Contents

Advertisement

Table 5-2:
Status alerts and Status Alert Severity
Alert code
Status message
A031
Low Power
A032
Meter Verification in Pro-
gress: Outputs to Fault
A033
Insufficient Right/Left Pick-
off Signal
A034
Meter Verification Failed
Meter Verification Aborted Fault
A035
A100
mA Output 1 Saturated
A101
mA Output 1 Fixed
A102
Drive Overrange
A104
Calibration in Progress
A105
Slug Flow
A106
Burst Mode Enabled
A107
Power Reset Occurred
A108
Basic Event 1 On
A109
Basic Event 2 On
A110
Frequency Output Satura-
ted
A111
Frequency Output Fixed
A112
Upgrade Transmitter Soft-
ware
A113
mA Output 2 Saturated
A114
mA Output 2 Fixed
A115
No External Input or Polled
Data
Configuration and Use Manual
(continued)
Default severity Notes
Fault
Applies only to flowmeters with the
enhanced core processor.
Varies
Applies only to transmitters with
Smart Meter Verification.
If outputs are set to Last Measured
Value, severity is Info. If outputs are
set to Fault, severity is Fault.
Fault
Applies only to flowmeters with the
enhanced core processor.
Fault
Applies only to transmitters with
Smart Meter Verification.
Applies only to transmitters with
Smart Meter Verification.
Informational
Can be set to either Informational or
Ignore, but cannot be set to Fault.
Informational
Can be set to either Informational or
Ignore, but cannot be set to Fault.
Informational
Informational
Can be set to either Informational or
Ignore, but cannot be set to Fault.
Informational
Informational
Can be set to either Informational or
Ignore, but cannot be set to Fault.
Informational
Normal transmitter behavior; oc-
curs after every power cycle.
Informational
Applies only to basic events.
Informational
Applies only to basic events.
Informational
Can be set to either Informational or
Ignore, but cannot be set to Fault.
Informational
Can be set to either Informational or
Ignore, but cannot be set to Fault.
Informational
Applies only to systems with trans-
mitter software earlier than v5.0.
Informational
Can be set to either Informational or
Ignore, but cannot be set to Fault.
Informational
Can be set to either Informational or
Ignore, but cannot be set to Fault.
Informational
Configure device options and preferences
Configurable?
No
No
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
55