Alert num-
ber
Alert title
A110
Frequency Output
Saturated
A111
Frequency Output
Fixed
A112
Upgrade Transmit-
ter Software
A113
mA Output 2 Satura-
ted
A114
mA Output 2 Fixed
A118
Discrete Output 1
Fixed
Configuration and Use Manual
Possible cause
The calculated Frequency Output is
outside the configured range.
The Frequency Output has been
configured to send a constant val-
ue.
The transmitter software is down-
level from the core processor soft-
ware.
The calculated mA Output value is
outside the configured range.
The mA Output is configured to
send a constant value.
The Discrete Output has been con-
figured to send a constant value.
Recommended actions
•
Check the Frequency Output scaling.
•
Check process conditions. Actual condi-
tions may be outside the normal condi-
tions for which the output is config-
ured.
•
Check for air in the flow tubes, tubes
not filled, foreign material in the tubes,
coating in the tubes, or other process
problems.
•
Verify that the measurement units are
configured correctly for your applica-
tion.
•
Purge the sensor tubes.
•
Cycling power to the meter or restart-
ing totalizers will restore the Frequency
Output to normal operation.
•
Check whether the output is in loop test
mode. If it is, unfix the output.
•
Check whether the output has been set
to a constant value via digital communi-
cation.
•
Contact customer support.
•
Check the settings of Upper Range Val-
ue and Lower Range Value.
•
Check process conditions. Actual condi-
tions may be outside the normal condi-
tions for which the output is config-
ured.
•
Check for air in the flow tubes, tubes
not filled, foreign material in the tubes,
coating in the tubes, or other process
problems.
•
Verify that the measurement units are
configured correctly for your applica-
tion.
•
Purge the sensor tubes.
•
Check whether the output is in loop test
mode. If it is, unfix the output.
•
Exit mA Output trim, if applicable.
•
Check whether the output has been set
to a constant value via digital communi-
cation.
•
Check whether the output is in loop test
mode. If it is, unfix the output.
Troubleshooting
129