Series 3700A System Switch/Multimeter Reference Manual

trigger.timer[N].EVENT_ID

This constant specifies the trigger timer event number.
Type
TSP-Link accessible
Constant
Yes
Usage
eventID = trigger.timer[N].EVENT_ID
eventID
N
Details
This constant is an identification number that identifies events generated by this timer.
Set the stimulus of any trigger object to the value of this constant to have the trigger object respond to
events from this timer.
Example
scan.trigger.channel.stimulus = trigger.timer[2].EVENT_ID
Also see
None

trigger.timer[N].overrun

This attribute indicates if an event was ignored because of the event detector state.
Type
TSP-Link accessible
Attribute (R)
Yes
Usage
overrun = trigger.timer[N].overrun
overrun
N
Details
This command indicates if an event was ignored because the event detector was already in the
detected state when the event occurred.
This is an indication of the state of the event detector built into the timer itself. It does not indicate if
an overrun occurred in any other part of the trigger model or in any other construct that is monitoring
the delay completion event. It also is not an indication of a delay overrun.
3700AS-901-01 Rev. D/June 2018
Affected by
The trigger event number
Trigger timer number (1 to 4)
Affected by
Instrument reset
Recall setup
Trigger timer N clear
Trigger timer N reset
Trigger overrun state (true or false)
Trigger timer number (1 to 4)