Section 11: TSP command reference
Details
This command saves the calibration constants and adjustment date and increases the adjustment
count by 1. The adjustment count is the number of times calibration has been saved.
This command does not check for errors in calibration data. Calibration data is saved regardless of
calibration data errors.
The calibration date can be specified with dmm.adjustment.date. If it is not specified, the date is
based on the system date.
To prevent data loss, you need to send the save command before locking calibration.
An error is generated if this command is issued when calibration is already locked.
Example
dmm.calibration.save()
Also see
dmm.adjustment.date
dmm.calibration.lock()

dmm.calibration.unlock()

dmm.calibration.unlock()
This function unlocks calibration if calibration was locked.
Type
TSP-Link accessible
Function
Yes
Usage
dmm.calibration.unlock("password")
password
Details
If the password does not match the saved password, an error is generated. The default password
from the factory is "KI003706". You can change the default with dmm.calibration.password.
Example
dmm.calibration.unlock("KI003706")
Also see
dmm.calibration.lock()
dmm.calibration.password
11-158
(on page 11-144)
(on page 11-156)
(on page 11-158)
Affected by
A string representing the password to unlock calibration
(on page 11-156)
(on page 11-157)
Series 3700A System Switch/Multimeter Reference Manual
Saves calibration data.
Where saved
Default value
Unlocks calibration using the default
password.
3700AS-901-01 Rev. D/June 2018