Series 3700A System Switch/Multimeter Reference Manual
Details
The trigger.timer[N].reset() function resets the following attributes to their factory defaults:
▪
trigger.timer[N].count
▪
trigger.timer[N].delay
▪
trigger.timer[N].delaylist
▪
trigger.timer[N].passthrough
▪
trigger.timer[N].stimulus
It also clears trigger.timer[N].overrun.
Example
trigger.timer[1].reset()
Also see
trigger.timer[N].count
trigger.timer[N].delay
trigger.timer[N].delaylist
trigger.timer[N].overrun
trigger.timer[N].passthrough
trigger.timer[N].stimulus
trigger.timer[N].stimulus
This attribute describes the event that starts the trigger timer.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
eventID = trigger.timer[N].stimulus
trigger.timer[N].stimulus = eventID
eventID
N
3700AS-901-01 Rev. D/June 2018
(on page 11-412)
(on page 11-413)
(on page 11-414)
(on page 11-415)
(on page 11-416)
(on page 11-417)
Affected by
Reset
Recall setup
Trigger timer N reset
The event that triggers the timer delay
Trigger timer number (1 to 4)
Section 11: TSP command reference
Resets the attributes associated with timer 1 back
to factory default values.
Where saved
Create configuration script
Save setup
Default value
0
11-417