Buffervar.seconds - Keithley 3700A Series Reference Manual

System switch/multimeter
Hide thumbs Also See for 3700A Series:
Table of Contents

Advertisement

Section 11: TSP command reference

bufferVar.seconds

This attribute contains the timestamp of a reading in seconds, in UTC format.
Type
TSP-Link accessible
Attribute (R)
Yes
Usage
seconds = bufferVar.seconds[N]
seconds
bufferVar
N
Details
The bufferVar.seconds information from a reading buffer is only available if the
bufferVar.collecttimestamps attribute is set to 1 (default setting). If it is set to 0, you will not
be able to access any time information from a reading buffer. You may change the collect timestamps
setting when the buffer is empty (bufferVar.clear()).
This attribute contains the nonfractional seconds portion of the timestamp when the reading was
stored in Coordinated Universal Time (UTC) format.
The nonfractional seconds portion of the timestamp gives the lowest resolution down to 1 second. To
access additional resolution of a timestamp, see bufferVar.fractionalseconds.
Example
reset()
testData = dmm.makebuffer(1000)
testData.collecttimestamps = 1
dmm.nplc = 0.5
dmm.range = 0
dmm.configure.set("Dcv_100mV")
dmm.setconfig("slot2", "Dcv_100mV")
scan.create("2035:2040")
scan.execute(testData)
print(testData.seconds[1])
printbuffer(1, 6, testData.seconds)
Also see
bufferVar.clear()
bufferVar.collecttimestamps
Reading buffers
11-32
Affected by
Buffer storage settings
Clearing the buffer
Instrument reset
Recall setup
The nonfractional seconds portion of the timestamp when the reading was stored
The reading buffer
The reading number N; can be any value from 1 to the number of readings in the
buffer; use the bufferVar.n command to determine the number of readings in the
buffer
(on page 11-21)
(on page 11-24)
(on page 6-7, on page 6-1)
Series 3700A System Switch/Multimeter Reference Manual
Where saved
USB flash drive using
dmm.savebuffer or
dmm.appendbuffer
This example creates a reading buffer
named testData, configures the buffer to
collect time and date data, sets and saves
the DMM configuration, creates a scan list,
and then runs the scan.
The print() command then outputs the
seconds portion of the timestamp of the first
reading in the reading buffer.
Output:
1.310375688e+09
The printbuffer() command then
outputs the seconds portion of the
timestamps for measurements 1 to 6 in the
reading buffer.
Output:
1.310375688e+09, 1.310375688e+09,
1.310375688e+09, 1.310375688e+09,
1.310375688e+09, 1.310375688e+09
3700AS-901-01 Rev. D/June 2018
Default value
Not applicable

Advertisement

Table of Contents
loading

Table of Contents