Dmm.nplc - Keithley 3700A Series Reference Manual

System switch/multimeter
Hide thumbs Also See for 3700A Series:
Table of Contents

Advertisement

Series 3700A System Switch/Multimeter Reference Manual

dmm.nplc

This attribute sets the integration rate in line cycles for the DMM for the function selected by dmm.func.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
value = dmm.nplc
dmm.nplc = value
value
Details
This attribute is not applicable for "frequency", "period", and "nofunction". If you query this
attribute for one of these functions, nil is returned. Note that "continuity" is fixed at 6.000000E-003
and cannot be changed.
The setting for NPLC may be adjusted based on what the DMM supports. Therefore, after setting the
NPLC, query the value to see if it was adjusted.
For dmm.nplc settings that are less than 0.2, sending dmm.AUTOZERO_ONCE results in significant
delays. For example, the delay time at an NPLC of 0.0005 is 2.75 s. The delay time at 0.199 is
5.45 s.
An error is generated if the command is used when dmm.func is set to "frequency", "period",
"continuity", or "nofunction".
The NPLC setting is saved with the dmm.func function setting, so if you use another function, then
return to the previous function, the NPLC setting you set previously is retained.
Example
dmm.func = "twowireohms"
dmm.nplc = 0.5
dmm.func = "dcvolts"
dmm.nplc = 0.1
Also see
dmm.aperture
dmm.autozero
dmm.configure.recall()
dmm.configure.set()
dmm.func
3700AS-901-01 Rev. D/June 2018
Affected by
Reset
DMM reset
Recall setup
The integration rate in line cycles:
60 Hz: 0.0005 to 15
50 Hz: 0.0005 to 12
(on page 11-145)
(on page 11-150)
(on page 11-165)
(on page 11-166)
(on page 11-179)
Section 11: TSP command reference
Where saved
Default value
Create configuration script
1.000000E+000
Save setup
Set the NPLC for 2-wire ohms to 0.5, then
set the NPLC for DC volts to 0.1.
11-209