Dmm.linesync - Keithley 3700A Series Reference Manual

System switch/multimeter
Hide thumbs Also See for 3700A Series:
Table of Contents

Advertisement

Series 3700A System Switch/Multimeter Reference Manual

dmm.linesync

This attribute determines if line synchronization is used during the measurement.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
state = dmm.linesync
dmm.linesync = state
state
Details
This attribute is only valid when dmm.func is set to "commonsideohms", "continuity",
"dccurrent", "dcvolts", "fourwireohms", "temperature", or "twowireohms". All other
functions generate an error when written and return nil when queried.
When line synchronization is enabled, measurements are initiated at the first positive-going zero
crossing of the power line cycle after the trigger.
The line sync setting is saved with the dmm.func function setting, so if you use another function, then
return to a previous function, the line sync you set previously is retained.
Example
dmm.func = "fourwireohms"
dmm.linesync = dmm.ON
Also see
Line cycle synchronization
dmm.configure.recall()
dmm.configure.set()
dmm.func
3700AS-901-01 Rev. D/June 2018
Affected by
Reset
DMM reset
Recall setup
Enable line sync: dmm.ON or 1
Disable line sync: dmm.OFF or 0
(on page 4-4)
(on page 11-165)
(on page 11-166)
(on page 11-179)
Section 11: TSP command reference
Where saved
Default value
Create configuration script
0 (dmm.OFF)
Save setup
Enables line sync for the "fourwireohms"
function.
11-195