Section 11: TSP command reference
Also see
Math calculations
dmm.configure.recall()
dmm.configure.set()
dmm.math.format

dmm.math.mxb.mfactor

dmm.math.mxb.mfactor
This attribute specifies the scale factor, m, for the y = mx + b math operation.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
scaleFactor = dmm.math.mxb.mfactor
dmm.math.mxb.mfactor = scaleFactor
scaleFactor
Details
This is not available for "nofunction". If command is queried when "nofunction" is selected, nil
is returned. If it is written when "nofunction" is selected, an error is returned.
This command sets the scale factor (m) for an mx + b operation for the selected measurement
function.
The mx + b math operation lets you manipulate normal display readings (x) mathematically according
to the following calculation:
y = mx + b
Where:
y is the displayed result
m is a user-defined constant for the scale factor
x is the measurement reading (if you are using a relative offset, this is the measurement with
relative offset applied)
b is the user-defined constant for the offset factor
Example
dmm.math.mxb.mfactor = 0.80
Also see
Math calculations
dmm.configure.recall()
dmm.configure.set()
dmm.math.format
dmm.math.mxb.bfactor
11-202
(on page 4-49)
(on page 11-165)
(on page 11-166)
(on page 11-200)
(on page 11-202)
Affected by
Reset
DMM reset
Recall setup
The scale factor; valid range is 4294967295 to 4294967295
(on page 4-49)
(on page 11-165)
(on page 11-166)
(on page 11-200)
(on page 11-201)
Series 3700A System Switch/Multimeter Reference Manual
Where saved
Create configuration script
Save setup
Sets the scale factor for the mx +b operation
to 0.80.
3700AS-901-01 Rev. D/June 2018
Default value
1.000000E+00