Section 11: TSP command reference

channel.connectsequential

This attribute controls whether or not channels are closed sequentially.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
sequential = channel.connectsequential
channel.connectsequential = sequential
sequential
Details
When channel.connectsequential is enabled, the list of channel actions is acted on
sequentially. No two relays are opened or closed simultaneously.
Using a sequential close allows you to determine the time for a close operation to happen. For
example, if you close three channels and each takes 4 ms to close (assuming no additional user
delay times), with sequential on, it will take 12 ms. With sequential off, it may be 4, 8 or 12 ms,
depending on whether or not the card can close multiple channels at once.
The order in which channels are opened or closed is not guaranteed with sequential off.
The sequential setting affects all channels in the instrument.
Example
channel.connectsequential = channel.ON
Also see
channel.connectrule
Switching and scanning
11-50
Affected by
Instrument reset
Recall setup
channel.OFF or 0: Disable sequential connections
channel.ON or 1: Enable sequential connections
(on page 11-48)
(on page 5-1)
Series 3700A System Switch/Multimeter Reference Manual
Where saved
Create configuration script
Specifies that channels close sequentially.
3700AS-901-01 Rev. D/June 2018
Default value
channel.OFF