Series 3700A System Switch/Multimeter Reference Manual

scan.scancount

This attribute sets the number of times the scan is repeated.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
scanCount = scan.scancount
scan.scancount = scanCount
scanCount
Details
The scan count attribute setting indicates how many times the scan list is iterated through before the
scan completes.
During a scan, the instrument iterates through the arm layer of the trigger model the specified number
of times. After performing the specified number of iterations, the instrument returns to an idle state.
Example
scan.create("1001:1009")
scan.scancount = 100
scan.scaninterval = 1.0 -- delay between scans
Create a scan that includes channels 1 to 9 of slot 1.
Set the scan count to 100.
Set the delay between scans to 1 s.
Also see
Scanning and triggering
Trigger model
3700AS-901-01 Rev. D/June 2018
Affected by
Reset
Scan reset
Recall setup
The scan count value (1 to 2,000,000,000)
(on page 5-23)
(on page 5-24)
Section 11: TSP command reference
Where saved
Default value
Create configuration script
1
Save setup
11-325