Scan.trigger.channel.stimulus - Keithley 3700A Series Reference Manual

System switch/multimeter
Hide thumbs Also See for 3700A Series:
Table of Contents

Advertisement

Series 3700A System Switch/Multimeter Reference Manual

scan.trigger.channel.stimulus

This attribute determines which trigger events cause the channel actions to occur.
Type
TSP-Link accessible
Attribute (RW) Yes
Usage
eventID = scan.trigger.channel.stimulus
scan.trigger.channel.stimulus = eventID
eventID
Details
This attribute selects which events cause the channel event detector to enter the detected state. Set
this attribute to 0 to start the channel action immediately at the default setting.
Set eventID to one of the existing trigger event IDs shown in the following table.
Trigger event IDs
Trigger event ID
channel.trigger[N].EVENT_ID or 41 to 48
digio.trigger[N].EVENT_ID or 1 to 14
display.trigger.EVENT_ID or 39
dmm.trigger.EVENT_LIMIT1_HIGH or 53
dmm.trigger.EVENT_LIMIT1_LOW or 52
dmm.trigger.EVENT_LIMIT2_HIGH or 55
dmm.trigger.EVENT_LIMIT2_LOW or 54
trigger.EVENT_ID or 40
trigger.blender[N].EVENT_ID or 58 to 59
trigger.timer[N].EVENT_ID or 20 to 23
tsplink.trigger[N].EVENT_ID or 17 to 19
lan.trigger[N].EVENT_ID or 29 to 36
scan.trigger.EVENT_SCAN_READY or 24
scan.trigger.EVENT_SCAN_START or 25
scan.trigger.EVENT_CHANNEL_READY or 28
scan.trigger.EVENT_MEASURE_COMP or 56
scan.trigger.EVENT_SEQUENCE_COMP or 50
scan.trigger.EVENT_SCAN_COMP or 26
scan.trigger.EVENT_IDLE or 27
schedule.alarm[N].EVENT_ID or 37 to 38
3700AS-901-01 Rev. D/June 2018
Affected by
Instrument reset
Recall setup
Scan reset
Trigger stimulus used for the channel action; see Details for possible trigger event IDs
Where saved
Default value
Create
50
configuration script
(scan.trigger.EVENT_SEQUENCE_COMP
Save setup
)
Description
The trigger event generated by the channel trigger N.
An edge (either rising, falling, or either based on the
configuration of the line) on the digital input line.
The trigger key (TRIG) on the front panel is pressed.
A DMM trigger event that indicates a measurement has
exceeded the high limit value on limit 1.
A DMM trigger event that indicates a measurement has
exceeded the low limit value on limit 1.
A DMM trigger event that indicates a measurement has
exceeded the high limit value on limit 2.
A DMM trigger event that indicates a measurement has
exceeded the low limit value on limit 2.
A *trg message on the active command interface. If
GPIB is the active command interface, a GET message
also generates this event.
A combination of events has occurred.
A delay expired.
An edge (either rising, falling, or either based on the
configuration of the line) on the TSP-Link trigger line.
A LAN trigger event has occurred.
Scan ready event.
Scan start event.
Channel ready event.
Measure complete event.
Sequence complete event.
Scan complete event.
Idle event.
Trigger event generated by the alarm N.
Section 11: TSP command reference
11-331

Advertisement

Table of Contents
loading

Table of Contents