Scan.trigger.measure.stimulus - Keithley 3700A Series Reference Manual

System switch/multimeter
Hide thumbs Also See for 3700A Series:
Table of Contents

Advertisement

Section 11: TSP command reference

scan.trigger.measure.stimulus

This attribute selects the trigger stimulus of the event detector trigger.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
eventID = scan.trigger.measure.stimulus
scan.trigger.measure.stimulus = eventID
eventID
Details
This attribute selects the events that cause the measurement event detector to enter the detected
state. You can use this to pace each one of the measurement count readings with an event.
To pace all readings by a single event, use scan.trigger.sequence.stimulus.
To bypass waiting for an event, set this to 0.
eventID can be set to one of the existing trigger event IDs, shown in the following table.
Trigger event IDs
Trigger event ID
channel.trigger[N].EVENT_ID or 41 to 48
digio.trigger[N].EVENT_ID or 1 to 14
display.trigger.EVENT_ID or 39
dmm.trigger.EVENT_LIMIT1_HIGH or 53
dmm.trigger.EVENT_LIMIT1_LOW or 52
dmm.trigger.EVENT_LIMIT2_HIGH or 55
dmm.trigger.EVENT_LIMIT2_LOW or 54
trigger.EVENT_ID or 40
trigger.blender[N].EVENT_ID or 58 to 59
trigger.timer[N].EVENT_ID or 20 to 23
tsplink.trigger[N].EVENT_ID or 17 to 19
lan.trigger[N].EVENT_ID or 29 to 36
11-334
Affected by
Instrument reset
Recall setup
Scan reset
The trigger stimulus that is used for the measurement event
Series 3700A System Switch/Multimeter Reference Manual
Where saved
Default value
Create
0
configuration script
Save setup
Description
The trigger event generated by the channel trigger N.
An edge (either rising, falling, or either based on the
configuration of the line) on the digital input line.
The trigger key (TRIG) on the front panel is pressed.
A DMM trigger event that indicates a measurement has
exceeded the high limit value on limit 1.
A DMM trigger event that indicates a measurement has
exceeded the low limit value on limit 1.
A DMM trigger event that indicates a measurement has
exceeded the high limit value on limit 2.
A DMM trigger event that indicates a measurement has
exceeded the low limit value on limit 2.
A *trg message on the active command interface. If
GPIB is the active command interface, a GET message
also generates this event.
A combination of events has occurred.
A delay expired.
An edge (either rising, falling, or either based on the
configuration of the line) on the TSP-Link trigger line.
A LAN trigger event has occurred.
3700AS-901-01 Rev. D/June 2018

Advertisement

Table of Contents
loading

Table of Contents