Series 3700A System Switch/Multimeter Reference Manual
scan.mode
This attribute sets the relay action when the scan starts.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
scanModeSetting = scan.mode
scan.mode = scanModeSetting
scanModeSetting
Details
When this attribute is set to open all, channels on all slots are opened before a scan starts.
When this attribute is set to scan.MODE_OPEN_SELECTIVE, an intelligent open is performed.
Assuming all steps being scanned have a function value of "nofunction" with their DMM
configuration then:
•
All channels and analog backplane relays used in scanning are opened
•
Closed channels and backplane relays not used in scanning remain closed during the scan
If any step has a DMM configuration with a function set to any other value than "nofunction":
•
Analog backplane relays 1 and 2 are opened on all slots
•
Any commonside ohms backplane relays are opened on all slots
•
Any amp channels are opened on all slots
•
All channels and backplane relays involved in scanning are opened
•
If a closed channel or backplane relay is not involved in scanning, it remains closed during the
scan
•
All channels are opened on any bank that contains backplane relays that are involved in scanning
When this attribute is set to scan.MODE_FIXED_ABR, it is equivalent to setting
MODE_OPEN_SELECTIVE, except:
•
All required backplane relays are closed before the start of the scan
•
These backplane relays are not opened or closed during the scan
•
These backplane relays do not open at the end of scan
Example
scan.mode = scan.MODE_OPEN_SELECTIVE
Also see
scan.reset()
Scanning and triggering
3700AS-901-01 Rev. D/June 2018
Affected by
Reset
Scan reset
Recall setup
The present scan mode setting. Set to one of the following values:
▪
scan.MODE_OPEN_ALL or 0
▪
scan.MODE_OPEN_SELECTIVE or 1: See Details
▪
scan.MODE_FIXED_ABR or 2: See Details
(on page 11-324)
(on page 5-23)