Preparing

Assembling

Assembling

Mounting the expiration valve
The parts must be entirely dry to prevent malfunctioning.
Hold stopper by the flange and place diaphragm on
the collar of the stopper.
Be careful to fit the diaphragm properly.
Insert stopper with diaphragm on top into the housing
from below and screw in tightly.
If the expiration valve has an optional water trap:
Fit the collecting jar.
Medicament nebuliser
Assemble in accordance with separate Instructions
for Use.
Install, see page 89.