Medicament nebuliser (option)
1 Remove the nebuliser hose from the nebuliser and
from the port on the device.
2 Remove the medicament nebuliser from the hose
system or
2 Dismantle the medicament nebuliser from the
paediatric hose system.
3 Remove the catheter connector (ISO cone ø15 / ø11)
from the inlet.
4 Remove adapter (ISO cone ø22 / ø11) from the
outlet.
5 Remove corrugated hose from the adapter.
Dismantle the medicament nebuliser in accordance
with its specific Instructions for Use.
Prepare the individual parts of the medicament
nebuliser and the adapting components for
disinfecting and cleaning in the autoclave.

Ventilation hoses

Remove from the adapters and ports.
Remove the water traps from the ventilation hoses.
Remove the collecting jars from the water traps.
Prepare the ventilation hoses, water traps and their
collecting jars and the Y-piece for cleaning in the
autoclave.

Flow sensor

Tilt the control unit upward.
6 Push the flow sensor to the left as far as it will go
and
7 pull out.
The flow sensor cannot be disinfected/cleaned by
autoclaving and cannot be sterilised by the hot steam
method.
Disinfect the flow sensor for about 1 hour in 70 %
ethanol solution.
Expose the sensor to air for at least 30 minutes.
Otherwise residual alcohol could damage the
sensor beyond control due to ignition during
calibration.
The flow sensor may be reused as long as calibration
can be carried out successfully.
2
1
4
5
2
6
Preparing
Dismantling
3
1
7
137