Evita V series / Babylog VN series
Batteries
Supplement
WARNING
To properly use this medical device, read and comply
with the instructions for use and this supplement.
Evita V series / Babylog VN series

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Evita V Series and is the answer not in the manual?

Questions and answers

Summary of Contents for Dräger Evita V Series

  • Page 1 Evita V series / Babylog VN series Batteries Supplement WARNING To properly use this medical device, read and comply with the instructions for use and this supplement. Evita V series / Babylog VN series...
  • Page 2: Getting Started

    Supplement to the instructions for use Supplement to the instructions for use Evita V series / Babylog VN series WARNING ► To properly use this medical device, read and comply with the instructions for use and this supplement. ● Keep this supplement with the instructions for use.
  • Page 3 The test steps performed were successful. The test failed. The test step failed. The test was not performed. The test was not completely performed. The test step was not performed. The test step is active. Supplement Evita V series / Babylog VN series...
  • Page 4: Operation

    PS500 power supply unit is displayed. For further information see: "Battery concept", page 14. Battery parameter field In addition to the charge status indicator, the Battery parameter field can be configured. Supplement Evita V series / Babylog VN series...
  • Page 5 The battery test in question can be started or canceled. The remaining runtime is displayed during the battery test. Observe the following information: "Maintenance", page 7. Observe the following information: "Battery test", page 8. Supplement Evita V series / Babylog VN series...
  • Page 6: Troubleshooting

    !!! 254 Battery depleted The remaining calculated Immediately connect the operating time of the battery device to the mains power is less than 5 minutes. supply. Supplement Evita V series / Babylog VN series...
  • Page 7: Battery Maintenance

    Specialized service per- depending on the sonnel results of the battery test Battery maintenance 7.2.1 Information on battery maintenance To achieve the maximum life span of the batteries, the following actions are required: Supplement Evita V series / Babylog VN series...
  • Page 8 The following table shows the expected typical operating time due to the ageing of a new battery without operation with an activated gas supply unit GS500. If the batteries do not correspond to the approximate operating time listed, replacement of the batteries is recommended. Supplement Evita V series / Babylog VN series...
  • Page 9 Prerequisites for the battery test while a patient is being ventilated: – The device is connected to the central gas supply system. – The device is connected to the mains power supply. Supplement Evita V series / Babylog VN series...
  • Page 10 – Battery replacement in xx months – Current operating time This value is indicated to the nearest 5 or 10 minutes depending on the battery used and based on the present power consumption of the device. Supplement Evita V series / Babylog VN series...
  • Page 11 For charging subsequent to storage For storage up to 6 months –15 to 25 °C (5 to 77 °F) For storage up to 3 months –15 to 40 °C (5 to 104 °F) Supplement Evita V series / Babylog VN series...
  • Page 12: Operating Characteristics

    F6.3H 250V IEC 60127-2/V (2 pcs.) venti- lation unit Batteries The battery runtime applies when the bat- teries are fully charged and new and venti- lation is typical. Low temperatures may reduce the battery runtime. Supplement Evita V series / Babylog VN series...
  • Page 13 F1 and F2 fuse at PS500 behind the F15A 80 V UL248 side part Capacity 24 Ah Voltage 24 V Electric current 0 to 15 A Battery runtime if mains power supply is not available Without GS500 240 minutes Supplement Evita V series / Babylog VN series...
  • Page 14: Principles Of Operation

    Battery charge indication on the screen The charge status indicator shows the available battery charge determined by the electrochemical processes. When the batteries are, e.g., fully charged, this state is indicated by a corresponding symbol. Supplement Evita V series / Babylog VN series...
  • Page 15 The percentage value refers to the battery charge, which entails a different operating time depending on the age of the battery, for example. Supplement Evita V series / Babylog VN series...
  • Page 16 PS500 is discharged first. If the mains power supply has not been re- established, the device switches over to the internal battery after the battery runtime of the PS500 has elapsed. The switch-over is indicated by the Internal battery activated alarm. Supplement Evita V series / Babylog VN series...
  • Page 17 When the device is fitted with the GS500 gas supply unit, the device calculates the time for the Battery depleted alarm allowing for the power consumption of a GS500, regardless of whether the GS500 is activated or not. Supplement Evita V series / Babylog VN series...
  • Page 18 This page has been left blank intentionally. Supplement Evita V series / Babylog VN series...
  • Page 19 This page has been left blank intentionally. Supplement Evita V series / Babylog VN series...
  • Page 20 Manufacturer Drägerwerk AG & Co. KGaA Moislinger Allee 53 – 55 23542 Lübeck Germany +49 451 8 82-0 www.draeger.com 9510319 – en © Drägerwerk AG & Co. KGaA Edition: 1 – 2019-11 Á9510319HÈ Dräger reserves the right to make modifications to the medical device without prior notice.